The board and its sketch

eeg_stream/eeg_stream.ino is the Arduino sketch flashed to the board. It reads two analog pins 250 times a second and prints each reading as a CSV line over USB serial. The bridge (apps/server) reads those lines. That is the entire contract between hardware and software.

It was checked against a live board on 2026-09-14. Read straight off the serial port at 115200 baud, the banner and the t_us,ch0,ch1 cadence matched this source exactly, so this file is what was flashed. It compiles for arduino:avr:uno in 2124 bytes of flash and 226 bytes of RAM (arduino-cli 1.5.1, AVR core 1.8.8).

What you need

PartNotes
Arduino UNO R3, or a pin-compatible 5 V boardThe rig here reports as arduino:avr:uno.
Olimex SHIELD-EKG-EMGOne shield = one differential channel.
An electrode cable for the shieldOlimex sells matching passive cables (snap / disposable-gel leads).
Disposable gel electrodes, or cup electrodes + conductive pasteNot ultrasound gel. It couples sound, not current, and the drift it causes looks like biology.
USB cableData-capable, not charge-only.
A laptop you can run on batterySee Safety. It also removes most mains hum.

Wiring

  1. Seat the shield on the Arduino headers. Check that every pin is in its socket and nothing is offset by a row.
  2. Plug the electrode cable into the shield's input connector.
  3. Leave the shield's jumpers at their single-shield defaults. This sketch reads the shield output on A0 (and also prints A1). If your shield's channel-select jumper routes it to another analog pin, change CH0_PIN in the sketch.
    • The jumper names and defaults are in Olimex's SHIELD-EKG-EMG user manual, and deliberately not repeated here. A copied jumper table that is wrong is worse than a pointer to the right one.
  4. Connect USB to the laptop.

ch1 (A1) is printed but not shown to be a second input.

  • On the one rig measured, with the leads floating, ch1 correlated +0.89 with ch0. That is what a floating pin picking up the same room looks like.
  • The cheapest check: touch or short one channel's leads and see which column responds (apps/server/RIG.md).

Flashing

With arduino-cli:

arduino-cli core install arduino:avr
arduino-cli board list                       # find the port: /dev/cu.usbmodemXXXX, /dev/ttyACM0, COM3
arduino-cli compile --fqbn arduino:avr:uno apps/olimex-shield/eeg_stream
arduino-cli upload  --fqbn arduino:avr:uno -p <port> apps/olimex-shield/eeg_stream

Or in the Arduino IDE: open eeg_stream/eeg_stream.ino, choose Arduino UNO and the port, and press Upload.

Stop the bridge first. Only one program can hold a serial port. If the workbench's bridge has it open, the upload fails with a busy-port error.

Check that it is streaming

arduino-cli monitor -p <port> -c baudrate=115200
# ArduinoEKG ready
# t_us,ch0,ch1
4004,523,458
8004,410,425
  • t_us should step about 4000 µs apart.
  • With nothing attached, values sit near mid-scale (~510).
  • Lines starting with # are comments, and the bridge skips them.
  • Opening the port resets the Arduino, so the first sample arrives about 1.6 s later.

The contract, and what breaks it silently

SettingValueIf you change it
BAUD115200The bridge must match (/open takes baud).
SAMPLE_RATE250 HzSilent. The bridge labels samples at its declared rate, so change NOMINAL_RATE_HZ in apps/server/rig.py too.
CSV columnst_us,ch0,ch1Silent. The bridge picks the channel by column index (channelCol, default 1 = ch0).
Banner text# ArduinoEKG readySafe. Only a human reads it.

The board's clock is not exact.

  • The sketch schedules samples off the board's own micros(), so t_us always agrees with itself and cannot reveal a clock error.
  • Against a computer's clock, one board ran at 249.83 Hz (about −680 ppm), and that figure moved by 41 ppm within a day.
  • The bridge measures the real rate against the host clock (apps/server/timebase.py). Any lock-in analysis must use that measured rate.
  • t_us is a 32-bit micros(), so it wraps every 71.6 minutes.

Versioning

This file is the version. If you change it:

  1. edit it here and commit;
  2. re-flash;
  3. repeat the streaming check above, and confirm the board shows the change before trusting a recording against it;
  4. add a line to the repository CHANGELOG.md saying what changed on the board.

Safety

This is a hobby amplifier, not a medically isolated device.

  • Run it from a battery-powered laptop unplugged from mains.
  • Never connect anyone with an implanted pacemaker or other active implant.

The full list is in the repository README, and in the workbench, which shows it before every connection.

Rendered from apps/olimex-shield/README.md.