The rig, as measured

Measured 2026-08-18, live, with the board attached. Everything here was probed, not assumed — the first attempt guessed the OpenEEG convention (57600) and got zero samples for eleven seconds.

Port/dev/cu.usbmodem1301 (2026-08-18); /dev/cu.usbmodem21301 (2026-09-15) — the suffix follows the USB socket, so never hardcode it
Baud115200 — not 57600. The bridge's default is already 115200; a wrong guess is what cost the first probe.
Firmware banner# ArduinoEKG ready
Wire formatt_us,ch0,ch1 CSV, # comment lines, two channels
Sample period4000 µs → 250.0 Hz by the board's own clock — which is circular. Against the host clock: 249.839 Hz, −644 ppm (2026-09-15, below).
ADC10-bit, resting near mid-scale (511.5)
Scale7.9 µV/count — NOT externally verified. Derived in an earlier internal app from the shield's built-in calibration square wave at its nominal ~250 µV peak-to-peak (≈ 31.6 counts; 250 / 31.6 = 7.9). The cal amplitude is the manufacturer's nominal figure; neither it nor this scale has been checked with an externally injected signal of known amplitude. The datasheet gain implies 1.72. An earlier note called 7.9 "independently recovered from the samples"; that check multiplied by 7.9 and read 7.9 back, so it was circular and is withdrawn. rig.UV_PER_COUNT_PROVENANCE; exported as the BIDS sidecar's ScalingProvenance.

2026-09-15 — validated with leads attached to the shield and nobody on them

Everything below came off /dev/cu.usbmodem21301: two raw captures read straight from the port before the bridge held it (61 s and 125 s), then the bridge and four signal-panel client pages on the same open port. No subject, no recording kept.

The sample rate is 249.839 Hz, not 250 — and it matters for lock-in, not for anything else

  • Board clock: t_us steps are 3988–4012 µs (the AVR's 4 µs micros() resolution), averaging exactly 4000. Zero dropped samples in 46,212 rows across both captures. This is what the old "measured off t_us" line was measuring — the board agreeing with itself.
  • Host clock: read in bulk (the OS buffer never exceeded 305 bytes, so the reader never fell behind), timestamped per chunk, and fitted to the lower envelope of latency — latency can only add delay, so its minimum is the clean line. Rate error −644 ppm (slow), scatter of the per-5-s minima 0.16 ms over 123 s. The earlier 61 s capture, read line by line, gave −599 ppm in the same direction; the bulk-read figure is the one to trust. It rests on the Mac's clock being good to a few ppm (NTP-disciplined) and on nothing else.
  • Mains is not a usable referee at this precision. At an assumed 250 Hz, the 60 Hz line read 60.0055 Hz in the first capture and 60.0489 Hz in the second, a few minutes apart. That 43 mHz swing is the grid, and it is larger than the error being measured.
  • What −644 ppm does, measured on the second capture through the bridge's own /extract:
windowlock-in at 60.000 Hzlock-in at 60.049 Hz (mains as the board sees it)
5 sp 2.6e-17, 24.4 dBp 2.6e-17, 24.4 dB
20 sp 3.4e-4, 10.5 dB, amplitude down to 17%p 3.0e-23, 29.5 dB
40 sp 0.76, −5.5 dB — "not detected"p 3.0e-15, 22.6 dB

A line 40 dB above the floor, carrying 94% of the power, reads as absent at 40 s. This is the motivating case for measuring the rate at all, now shown on this rig: a 40 Hz drive played from the computer lands at ~40.026 Hz in these samples, and a minutes-long lock-in at the nominal frequency averages it toward zero. The extraction layer did not lie about it — it tiered the result MEASURED_UNSYNCED and said "bounds the response, does not prove absence" — but a reader who trusts the p-value would conclude the opposite of the truth. Any lock-in here needs a co-recorded reference or the measured rate. Ordinary displays, contact checks and heart rate are unaffected (a 0.06% timebase error is invisible to them).

  • bridge /status reported rate: 250, a constant, and consumers labelled it "measured" — one client's log said "Analysis uses the measured 250 Hz". Fixed the same day: see the next section.

The bridge now measures the rate itself (timebase.py)

  • The method above runs inside the daemon, continuously: the reader timestamps each chunk right after read() returns, RateEstimator bins latency minima every 5 s over a sliding 600 s window, and /status carries rateMeasured, rateErrorPpm, rateUncertaintyPpm and rateBasis. null until 30 s over 6 bins — never a guess. rate stays 250 and is now documented as declared.
  • The board drifts, so no constant could have held it. −644 ppm in the morning, −685 ppm that afternoon, −682.1 ± 0.4 ppm over 414 s at the end of the day. The afternoon figure was checked two ways on one capture: the offline method gave −685.1, the estimator −685.0 — the 41 ppm is the board, not the estimator.
  • Stale lines found a real bias. A capture taken right after another process released the port started with lines still in the OS buffer, delivered instantly, then 1.62 s of silence while the Arduino reset. That early instant became a bin minimum and dragged −685 ppm to −3300 ± 1491. Any 1 s gap now restarts the fit (GAP_RESET_S); the same data then reads −690.8 ± 2.3.
  • Who uses it: /extract, /quality, /bands and /spectrum with source:"live" use the measured rate once it exists and say which in rate_basis; a caller-supplied rate always wins. signal-panel exposes it as analysisRate, and its lock-in widget prints measured or DECLARED — not yet measured. Client captures record rate_hz and rate_basis.
  • What it recovers, live: 60 Hz at 40 s — p 0.86 at the declared 250 Hz, p 2.2e-4 at the measured rate. Through the panel on another client page, the lock-in widget found mains at p 1.6e-14 over 20 s against 249.830 Hz measured.
  • Assumes no samples lost on the wire (zero seen in 46,212 rows) and a host clock good to a few ppm. It cannot check either.

Mains, noise and the two channels

  • With leads on the shield and nothing on them: ch0 16.9 µV RMS (sd 2.1 counts), 94% of its power in 58–62 Hz, the 60.02 Hz line 40 dB above the median bin. Compare 2026-08-18 with nothing attached: 5.0 µV, span 3 counts. Floating leads are an antenna, as expected; the front end's nominal 40 Hz low-pass does not stop it. quality.assess returns unusable (amplitude far below the 100 µV scalp-EEG plausibility floor, mains 30–66× the 2–40 Hz power) — the correct answer.
  • 13 distinct ADC values in 61 s: the 7.9 µV step (itself the unverified, cal-derived scale) is half the noise — the dither boundary.
  • The analog low-pass cannot be seen in this data. Median power 45–55 Hz sits only 3.9 dB under 5–35 Hz because both are at the white quantisation floor; a leads-off capture measures the ADC, not the filter.
  • ch0 and ch1 correlate at +0.89 — +0.93 in the mains band, and still +0.64 with 55–65 Hz removed — and ch1 lags ch0 by 23° at 60 Hz (~1.1 ms), far more than the ~0.1 ms between the two back-to-back analogRead calls. ch1 is carrying the same environment through a different (higher-impedance) path. That is consistent with a floating pin and does not establish a second differential input; the touch test above is still the only thing that will.

The bridge and the panel on real hardware

  • /ports lists the device; /open returns in 30 ms, but the first sample arrives 1.62 s later — opening the port resets the Arduino, which then boots. A second /open while streaming kept total and generation (no reset), as designed.
  • The panel's "already-open port" fix holds on the real device: with the bridge holding 10,022 and then 27,092 samples, two client pages (an ECG monitor and a plant-signal reader) each received a first batch of exactly 500 samples (2 s), not the ring.
  • Every consumer badged the source MEASURED (status.demo false) and reported contact unusable; the ECG page's 16.8 µV / 29.57× matched the raw capture's 16.9 µV / 29.53×; the plant reader's config removed the alpha-ratio reason that the scalp defaults raise (the check is now readout-aware and does this itself for non-EEG readouts); both hookup guides gated the real port; a third client found the device without the no-device prompt and took a 3 s probe of 753 samples. The port was released at the end.

From the firmware, not measured

t_us is micros() in a 32-bit unsigned long, so it wraps to 0 every 71.6 minutes. The sketch's scheduling survives the wrap; anything that reads timing out of the t_us column across a longer session must handle it. Since 2026-09-25 the bridge reads t_us too (timebase.StreamIntegrity): it counts gaps (steps > 1.5 × the declared period) with modular arithmetic, so the wrap is a normal step, and a backwards jump is counted as a clock reset. It writes t_us into the black box and never inserts a sample.

What a no-electrode capture looks like here

With nothing attached (2500 samples, 10 s):

  • span 3 counts, amplitude 5.0 µV — the channel is at its own floor, as it should be
  • 60 Hz mains at +32.2 dB, p = 1.9e-26 — a decisive detection, and the single best proof the whole chain works: a real external signal is being picked up and correctly measured
  • 50 Hz not significant (p = 0.071), which is right for a 60 Hz country
  • in-band noise at 5.4 Hz: 322 nV over 10 s, inside the 64–750 nV range this tool's own validation found across six prior captures — though that range is at 40 Hz over 60 s, so the comparison is loose (range and count corrected 2026-09-13 from 65–750 across eight)

quality.assess returns unusable, and that is the correct answer: nothing is attached to a body. The verdict is about signal plausibility, not about the instrument, and it does not measure impedance.

Two channels, and what that does NOT yet establish

The sketch streams ch0 and ch1, and both move independently (sd 0.65 and 0.51 counts, means 509.4 and 502.8). This does not establish that the shield drives two independent differential inputs. A single SHIELD-EKG-EMG is one channel; ch1 may simply be a floating Arduino pin, and a floating pin also wanders by a fraction of a count.

It matters because any bilateral electrical recording design depends on it: if ch1 is not a second input, a second shield is needed. The cheapest test: short or touch one channel's inputs and see which column responds. Until that is run, treat ch1 as unidentified rather than as a second channel.

Consuming the stream: two traps, both hit here

There is no /recent HTTP route. recent() is a method on eeg-client.js, which keeps its own ring fed by /stream. A consumer that fetches /bridge/recent gets a 404 body where it expected samples, and — because the body is JSON — nothing throws.

A JSON proxy cannot carry /stream. urlopen().read() never returns on an SSE response. Any project proxying the bridge same-origin needs a line-by-line pipe; the workbench's /bridge/stream (apps/web/serve.py) is one.

And "open" is not "streaming." Reopening the port resets total to 0 and bumps generation, so a readiness check must watch for the counter to increase between polls rather than compare against a count taken before connecting.